Power Planning for Home Medical Equipment in Hollywood: Oxygen, CPAP and Outages
Oxygen concentrators, CPAP machines and other home medical devices need an outage plan. Circuits, battery backup, generators and transfer switches explained.

For many families in Hollywood, a power outage is an inconvenience. The lights go out, the food in the fridge needs watching, and everyone waits for FPL. For a family member who depends on an oxygen concentrator, a CPAP or BiPAP machine, a powered hospital bed, a lift chair, a suction machine or another electrically powered medical device, an outage is something else. It is a safety issue that deserves a plan before it happens.
This guide covers the electrical side of that plan: how to understand what your equipment needs, how to power it safely in your home, and what backup options make sense, from battery units to generators with a proper transfer switch. It is written from an electrician's point of view. It is not medical advice. Every decision about the equipment itself, including how long someone can safely be without it, belongs to the doctor and the equipment supplier.
If anyone's breathing or health is at risk during an outage, call 911.
Step 1: Know exactly what the equipment needs
Before you can plan for backup power, you need to know how much power each device uses and how it plugs in.
Read the label
Almost every medical device has a label or a nameplate on the back or bottom, and the same information is in the user manual. Look for:
- Volts (V): in a home, this is usually 120 volts.
- Amps (A) or Watts (W): how much power the device draws. If the label shows amps, you can estimate watts by multiplying amps by volts.
- Plug type: a standard two or three prong plug, or something special.
- Battery: whether the device has its own internal or external battery, and how long it lasts.
Write this down for every device. Include things that support the device, such as a humidifier attached to a CPAP or a heated tube.
Ask the supplier about backup
The company that supplies the equipment, often called a durable medical equipment supplier, can tell you:
- Whether there is a battery option for the device.
- Whether there is a backup that does not need power at all, such as oxygen cylinders.
- Whether the device is safe to run on a generator or a battery inverter.
- What they recommend for outages.
That conversation should come first. The electrical plan supports the medical plan, not the other way around.
Step 2: Make the everyday power setup safe
Many problems with home medical equipment have nothing to do with outages. They come from how the device is plugged in every day.
Plug directly into a wall outlet
Most manufacturers tell you to plug medical devices directly into a wall outlet, not into an extension cord or a power strip. A cheap cord or strip can overheat, especially with a device that runs for hours. Our article on extension cords and power strips explains why.
Use an outlet in good condition
A loose outlet that lets the plug wiggle out, or one that feels warm, is a problem. If an outlet has scorch marks, buzzing or a burning smell, stop using it and call an electrician. See what to do about a burning smell from an outlet.
Consider a dedicated circuit
If a medical device shares a circuit with a space heater, a hair dryer, a window AC unit or a microwave, the breaker can trip when everything runs at once. When that happens in the middle of the night, it is easy to miss. An electrician can check which outlets share a circuit and, if needed, add a dedicated circuit for the bedroom where the equipment runs.
Think about surge protection
Lightning is common in South Florida, and power can surge when it comes back after an outage. A whole home surge protector at the panel, plus a quality point of use surge protector where the manufacturer allows it, helps protect sensitive electronics. Our guide on surge protectors versus power strips covers the difference.
Keep oxygen away from heat and flames
If someone in the home uses oxygen, follow the supplier's safety rules, including keeping the equipment away from open flames, candles, gas stoves and heat sources. That matters even more during an outage, when people reach for candles. Use battery lanterns and flashlights instead.
Step 3: Choose a backup power approach
There is no single right answer. The best choice depends on the device, how long it needs to run, your home, and your budget. Most families end up with a combination.
Option A: The device's own battery
Some devices, such as many portable oxygen concentrators and some CPAP machines, can run on a dedicated battery made by the manufacturer. This is usually the simplest and safest backup because it was designed for that device.
Good for: short outages, moving to another room, or traveling to a shelter or a relative's home. Limits: runtime is limited, and batteries need to stay charged.
Option B: A battery backup unit or portable power station
A battery backup unit, sometimes called a UPS, or a larger portable power station can run a device for a while during an outage. These plug into the wall to charge and switch over when the power drops.
To estimate runtime, compare the device's wattage to the battery's capacity, which is often listed in watt hours. A rough estimate is the battery's watt hours divided by the device's watts, minus some loss. Real runtime is usually lower than the simple math suggests, so plan with a margin. Ask the manufacturer whether the device is approved for use with an inverter or a backup unit.
Good for: bridging the gap until power returns or a generator starts. Limits: larger units are heavy and need charging, and capacity runs out.
Option C: A portable generator, connected safely
A portable generator can run medical equipment for longer outages, but it has to be used correctly.
- Never run a generator indoors, in a garage, on a covered porch or near windows and doors. Generator exhaust contains carbon monoxide, which can kill. Place it well away from the house, and install battery powered carbon monoxide alarms inside. See our article on carbon monoxide alarms during generator season.
- Never plug a generator into a wall outlet to power the house. This is called backfeeding. It can electrocute utility workers and neighbors and damage your home. Read why backfeeding a generator is dangerous.
- Use a transfer switch or an approved interlock with an inlet, installed by an electrician, if you want the generator to power circuits inside the house.
- Or run a heavy duty, outdoor rated cord directly from the generator to the device, if the manufacturer allows it and the cord is rated for the load.
Some sensitive electronics prefer cleaner power. Many inverter style portable generators produce power that is suitable for electronics, but check with the device manufacturer.
Good for: longer outages. Limits: fuel storage, noise, maintenance, and someone has to start it.
Option D: A standby generator with an automatic transfer switch
A permanently installed standby generator, connected through an automatic transfer switch, starts on its own when the power goes out and switches selected circuits, or the whole home, over to generator power. For a household where someone depends on equipment overnight, the automatic start can be a big advantage, because nobody has to wake up and do anything.
This is a larger project. It usually involves a permit, a fuel supply such as natural gas or propane, a concrete pad, and careful sizing so the generator can handle the medical equipment plus the other circuits you choose. Learn more about automatic transfer switches and how to think about generator size.
Good for: households that need hands free backup for long outages. Limits: cost, space and maintenance.
Option E: A home battery system
Some homes use a whole home or partial home battery system, sometimes paired with solar. These can switch over automatically and quietly. Capacity varies widely, so sizing matters. See our overview of home battery backup systems.
Talk to an Electrician
We answer the phone fast
From panels and wiring to lighting, generators, EV chargers and three phase power, we cover South Florida homes and businesses, including emergency repairs. Share the problem and we map out the repair.
Call (954) 602-0050A worked example, with made up numbers
Here is how the math works, using round numbers that are only an illustration. Use the real figures from your own device label and your own battery.
Say a device label lists a draw of 1.5 amps at 120 volts. Multiply the two and you get about 180 watts. That is the most it might draw. Many devices draw less most of the time, but planning with the label figure keeps you on the safe side.
Now say you are looking at a battery unit rated at 1,000 watt hours. Divide 1,000 by 180 and you get a little over 5 hours. Inverters lose some energy as they convert battery power, and batteries rarely deliver their full rating, so plan for noticeably less, perhaps 4 hours in this example. If the outage could last longer than that, the battery is a bridge, not the whole plan.
Repeat the math for anything else you want to run from the same battery, such as a fan or a phone charger, and add the watts together before you divide.
Questions to ask your electrician
When you talk to an electrician about backup power for medical equipment, these questions help you compare options:
- Which outlets in the bedroom share a circuit with other loads?
- Is my panel in good condition, and does it have room for a new circuit or an interlock?
- What size generator or battery would cover the circuits on my priority list?
- Would you recommend a manual transfer switch, an interlock with an inlet, or an automatic transfer switch for my home?
- Does the work need a permit, and will you handle it?
- How do I test the system once it is installed?
A good electrician will explain the trade offs in plain language and will never pressure you into a larger system than you need.
Step 4: Decide which circuits matter most
If you go with a transfer switch, an interlock or a standby generator, you choose which circuits get backup power. For a family with medical equipment, a typical priority list looks like this:
- The bedroom circuit where the medical device runs.
- The refrigerator, especially if any medication needs to stay cold.
- A few lights and the outlet where phones charge.
- A fan or, if the generator is large enough, part of the air conditioning. South Florida heat can be a health risk on its own.
An electrician can map your panel, label the circuits clearly, and help you decide what makes sense for the generator or battery you choose.
Step 5: Register and tell the right people
Some steps on the list are not electrical, but they matter.
Your utility
FPL runs a Medically Essential Service program for residential customers whose electric service is medically essential, meaning someone in the home depends on electrically powered equipment to avoid loss of life or serious complications that would put them in the hospital. The application has two parts: the customer fills out one part, and a physician licensed in Florida completes a certificate that describes the equipment and why it is needed. Enrollment does not renew on its own, so put a reminder on the calendar to reapply every year.
Be clear about what the program does and does not do. Enrolled customers get special notices, including notices before and after a hurricane. FPL states plainly that the program does not guarantee uninterrupted service and does not give priority restoration after a storm. In other words, it is a communication tool, not a backup plan. You are still responsible for backup equipment and power, which is exactly why the rest of this guide matters.
Your county
Broward County runs a Special Needs Shelter program for residents whose medical needs are more than a general population shelter can handle but who do not need a hospital. People who depend on electricity for medical equipment are specifically told to plan ahead. You pre register through the Florida Special Needs Registry website by choosing Broward County, or you can ask Broward County Emergency Management to mail you a paper application. Pre registration is open all year, it is not required, but the county strongly encourages it so the shelter is ready for you, and registrants are asked to submit a new application every two years.
Treat the shelter as a backup to your own plan, not the plan itself. The county describes these shelters as having backup electricity for limited lighting, air conditioning and essential medical equipment, staffed by medical personnel, but without the comfort of home. If you can stay with family or friends outside the outage area, that is usually the better choice. Do the paperwork well before storm season, because a physician's information is part of the application and reviews take time.
Your household and neighbors
Make sure everyone in the home knows where the backup batteries, flashlights and generator are, how to start them, and who to call. A neighbor who knows your situation can check in during a long outage.
Step 6: Practice and maintain
A backup plan only works if it works on the day you need it.
- Test batteries every month or two. Charge them fully and note the date.
- Run the generator under load on a regular schedule, following the manufacturer's guidance. Keep fuel fresh and stored safely outside the house. See our standby generator maintenance schedule.
- Practice a switchover once, in daylight, so it is not the first time during a storm at night.
- Keep a printed sheet near the equipment with the device's power needs, the supplier's phone number, the doctor's phone number, the utility outage line and your electrician's number.
A simple outage day plan
When the power goes out:
- Make sure the person using the equipment is safe and the device is running on its battery or backup.
- Check whether the outage is just your home. If the neighbors have power, check your main breaker. If it keeps tripping, do not keep resetting it, and call an electrician.
- Report the outage to FPL.
- If you have a portable generator, set it up outdoors, far from windows and doors, and connect it the safe way.
- Keep phones charged and stay in touch with your supplier if the outage looks long.
- If the backup will not last and power is not coming back, follow your plan to move to a relative's home, a hospital or a shelter, as your doctor advised.
Hollywood homes and older electrical systems
Many homes in Hollywood were built decades ago, and some still have older panels, limited circuits or wiring that was not designed for today's loads. If you are adding medical equipment, a generator inlet or a battery system, it is a good moment to have the panel and wiring checked. We often find shared circuits, loose connections or outdated panels that are worth fixing before anything else is added. See our page on Federal Pacific and Zinsco panels in Hollywood.
Frequently asked questions
Can I plug an oxygen concentrator into a power strip?
Many manufacturers tell you not to. Follow the device manual, and plug directly into a wall outlet in good condition whenever possible.
Will a regular portable generator damage my CPAP?
Some electronics prefer cleaner power. Many inverter generators are suitable for electronics, but check with the device manufacturer before you rely on any generator.
How long will a battery backup run my device?
Divide the battery's watt hours by the device's watts for a rough estimate, then plan for less. Ask the manufacturer for real runtime figures.
Do I need a permit for a transfer switch or standby generator?
Permanent installations generally need a permit and inspection. Your electrician should handle that.
Will FPL restore my power first if I am registered?
Utility programs for medical equipment users help with communication, but they generally do not guarantee priority restoration. Plan for your own backup.
What is the first thing to do if the power goes out at night?
Make sure the person using the equipment is safe and on backup power. If anyone is in danger, call 911.
Get a power plan for your home medical equipment
24/7 Electrician helps families in Hollywood, Dania Beach, Hallandale Beach, Davie, Pembroke Pines and nearby cities plan safe power for medical equipment: dedicated circuits, surge protection, generator inlets, transfer switches and standby systems. Call (954) 602-0050 to talk through your situation.
Request Service
Tell us what is happening and we will take it from there
Use this form to give the office the full picture in one step: what is happening, the property location, entry details and the time you need help. Whoever calls you back has read it first.
If the power is out or something smells like it is burning, do not type. Call (954) 602-0050. We answer the phone fast.
Where We Work
36 cities across three counties
Every city has its own page for planned work, and its own page for after hours. Please choose the name of your city, or tap Emergency for urgent dispatch.
Broward County
- Coconut CreekEmergency
- Cooper CityEmergency
- Coral SpringsEmergency
- Dania BeachEmergency
- DavieEmergency
- Deerfield BeachEmergency
- Fort LauderdaleEmergency
- Hallandale BeachEmergency
- Hillsboro BeachEmergency
- HollywoodEmergency
- Lauderdale-By-The-SeaEmergency
- Lauderdale LakesEmergency
- LauderhillEmergency
- Lighthouse PointEmergency
- MargateEmergency
- MiramarEmergency
- North LauderdaleEmergency
- Oakland ParkEmergency
- ParklandEmergency
- Pembroke ParkEmergency
- Pembroke PinesEmergency
- PlantationEmergency
- Pompano BeachEmergency
- Southwest RanchesEmergency
- SunriseEmergency
- TamaracEmergency
- West ParkEmergency
- WestonEmergency
- Wilton ManorsEmergency